Resolution up to 0.01 Lux makes it possible for detecting dim light and weak light.
1. Maximum display value: 3999 2. Measuring range: 39.99, 399.9, 3999, 39990 3. Accuracy: ± 4% + 5 (in the case of 2856K standard light source) 4. V (λ) matching error: ƒ’1≤6% 5. Maximum resolution: 0.01 Lux 6. Sensor: silicon photodiode and visual correction filter 7. Repetitive accuracy: ± 2% 8. Sampling rate: 2 times/sec 9. Operating temperature and humidity: 0℃ - 40℃ (32℉ - 104℉) 0 - 70% RH 10. Storage temperature and humidity: -10℃ - 50℃ (14℉ - 122℉) 0 - 80% RH 11. Working power: single 6F22 9V battery 12. Service life of battery: about 200 hours of continuous use 13. Meter size: host 160 x 78 x 43 mm Probe φ40 x 35 mm and 1000 mm shielding extension line 14. Meter weight: about 370 g (including battery and probe) 15. Deviation characteristic of cosine angle: 30° ± 2%; 60° ± 6%; 80° ± 25% |